Important Note
The home position calibration can be performed only when the parallelism is achieved. Refer to Einstein Pro XL: Check Parallelism and Calibration for more details.
- Remove the build platform from the printer and place it aside on a clean work surface.
- On the home screen, select Settings > Move & Calibration > Home Calibration > Manual Homing. The following screen appears.

- Select the material tray from the list. The following screen appears.

- Select Move to Home. The build platform holder moves down the Z-axis to the current home position.
- Set the value of a step by pressing the - and + icons of Step Value. This is the increment used for the Motion Distance. The Step Value increments can be set to 0.01, 0.10, 1.00, and 10.00.
- Set a motion distance of the platform by pressing the - and + icons of Motion Distance. This is the total distance the platform moves.
Tip: The positive values move the platform up; the negative values move the platform down. Pressing the refresh icon resets the motion distance value to zero.
- The adjustment should be made based on the average height of the calibration blocks.
- If the average height measurement is below 4.90 mm, move the platform up by selecting Move the required number of steps.
- If the average height measurement is above 5.10 mm, move the platform down by selecting Move the required number of steps.
- Set the motion distance to the required adjustment. Select Move to make the adjustment and then select Save.
- The build platform holder starts moving up to the top of the Z-axis. The new home position is set.