ProCure: Door Replacement

Learn how to safely remove and replace the ProCure 1 curing chamber door.

Overview

If the ProCure door becomes damaged, loose, or misaligned, it may need to be replaced. This article outlines the steps to remove the existing door and install a new one properly.


Before You Begin

  • Power off the unit.
  • Unplug the power cord.
  • Ensure the curing chamber is empty and cool to the touch.

Remove the Door Hinge Pin

  1. Open the curing chamber door fully.
  2. Locate the hinge pin connecting the door to the unit.
  3. Using needle-nose pliers, gently pull the hinge pin upward and out of the hinge assembly.
  4. Carefully lift the door away from the unit once the pin is removed.


Install the New Door

  1. Position the new door in alignment with the hinge brackets.
  2. Press firmly to ensure the hinge openings are properly aligned.
  3. Slide the hinge pin downward through the hinge assembly until fully seated.

Confirm the door opens and closes smoothly and sits flush against the curing chamber before returning the unit to operation.